Bridge Cover

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable15 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Non ferrous Standard Tool1---
Table 3. Safety

Procedure

  1. Undock patient table, and position table away from the magnet.
  2. Remove the bridge cover by removing two M6 bolts.
    Figure 1. Bridge Cover Removal
  3. Installation is reverse order of removal.

Finalization

No finalization steps.