Revision History
| Revision | Date | Reason For Change |
|---|---|---|
| 1 | Nov 16, 2018 | Initial Release |
| 2 | Oct 21, 2019 |
Chapter: Installation=>Guided Mechniacal Install Flow=>2rd day:cable connection. Correct the diagram of "Install Temperature & Humidity Sensor on GOC" run1321 to 1383. Add "Installing T5820 GOC" procedure. For "Installation of Fiber Optic Cables" add T5820 configurtion procedure and this procedure is same with T5810. Add T5820 cable map. Add GOCAA cable connection. Chapter:Installation=>Install and Calibration Wizard for R/RD Series Magnet In the calibration guided flow, figure 10 6th day, update "Sys Gain Cals" to "Sys Gain Cals (only for software SV25.2 and before system)" . In the calibration guided flow, figure 10 6th day, click into the link of "SPT Full Test", in section 3 "Invoking SPT Full Test Mode", step 2, add (For SV25.2 and before system),.... (For SV25.3 and above system), click start SPT to begin, the following items "Setup check" "EC Check" "SNR/Sysgain""Stability" "CNT" will be proceed automatically. Chapter:Adjust=>Cal=>System Level Procedures=>LVShim Procedures For SN2270524. Add words: For the software SV25.2 and above System, The value of grad shim X/Y/Z will looks bigger than before because the resolution for Gradshim values changed (shift by 4 bits). Chapter: Sys Diagrams=>System Block Diagram Add ICN7 cabinet diagram Chapter: Software=>Insite Remote Service Platform (RSvP) theory of operation New create topic group Chapter: Software=>System Host Software=>Loading host system software Update "Guided install Site Configuration Options" list. In system configuration, add items "Hospital name; Suite ID; Host ID; Unique / System ID; Service ID", remove item "Enable HO Shim"; In Hardware configuration, add items "Scan range; Wireless PAC Frequency Channel", update "Scanner Channel configuration: 8 for Creator, 16 for Explorer"; Change "HEC Type/Cooling Cabinet Type" to "Cooling Cabinet Type"; Move "UPM Type" under "Table Configuration"; Change "BRM/BRM2" to "BRM,NOTE: BRM stands for BRMx coils"; Add below items: "Detachable Table: This is for motors table only, not applicable for SIGNA Creator and SIGNA Explorer." "Number of Tables: This is for motors table only, not applicable for SIGNA Creator and SIGNA Explorer." "Table Encoder Calibration Value: Not applicable for SIGNA Creator and SIGNA Explorer."" Encoder Calibration value 2: Not applicable for SIGNA Creator and SIGNA Explorer." "Altitude in meters (0 to 5000): Not applicable for SIGNA Creator and SIGNA Explorer." update the pic "Sample of software revision to install" by removing advanced mode. Chapter: Software=>System Host Software=>Loading USB host system software (Quick Start) New create topic Chapter: Theory=>RF Subsystem=>Receive Chain Theory Update the theory diagram for ICN7 Chapter: Troubleshooting=>Diagnostics=>System Function=>Recon=>ICN Quick Health Check=>ICN Hardware Health for Dell ICN Add R640 ("GEN7") in the senstance "The ICN Hardware Health Diagnostic is used on the Dell R620 ("Gen 5") and R630 ("Gen 6") ICNs only" Chapter: Troubleshooting=>Diagnostics=>System Function=>Recon=>ICN Hardware Diagnostics=>ICN Hardware Intensive Diagnostics Add R640 ("GEN7") in the senstance "The ICN Hardware Health Diagnostic is used on the Dell R620 ("Gen 5") and R630 ("Gen 6") ICNs only" Chapter: Parts=>Illustrated parts=>PATIENT TRANSPORT LITE TABLE For SN2302216. Update "Straps and Pads for Patient Table" FRU list Chapter: Parts=>Illustrated Parts=>Low Hight Fixed Patient Table Add "5458804-2" for Table Control Box For SIMs feedback ID2594: update 5420619 description to "RF Cable FRU kit (RF Cable + Cable Clamp Assy)" For SN2302216. Update "Straps and Pads for Patient Table" FRU list. Update all the FRU list and structure. Chapter: Parts=>Illustrated parts=>OC Add T5820 PC FRU list. Chapter: Parts=>Illustrated parts=>System Cabinet Add GEN7 ICN in item13 and add table13. Add GP4 new FRU part 6237000-3 Chapter: Replacement=>Operator WorkSpace Replacement=>Dell T5820 computer replacement and Low Level FRU replacement New create topic group Chapter: Replacement=>System Cabinet=>ICN (Dell R640) New topic group including "ICN Replacement", "ICN SSD Replacement(Dell R-640)" "ICN Power Supply Replacement(Dell R640)" "ICN On/Off procedure for Dell R640(Gen 7)" Chapter: Replacement=>Low Height Fixed Table=>Table Control Box Add a note "If Table Control Box 5458804 installed in site need to be replaced, please order 5458804/5458804-2 for replacement. If Table Control Box 5458804-2 installed in site need to be replaced, please order 5458804-2 for replacement." Chapter: Replacement=>Coil Blower and Chiller Assemblies=>11kw Chiller Controller Replacement For SN2194811. New create topic. Chapter: Planned Maintenance Update PM list. |
| 3 | Sep 23th, 2020 |
SNDOC2312335 Update PM lists. SNDOC2325945 Update Coil FRUs SNDOC2302216 Patient handling accessories (pads straps) FRU introduction SNDOC2326404 Update LHFT FRU list and structure SNDOC2323031 Add 5376361 & 5376361-2 encoder box FRU SNDOC2296770 Add 6237000-3 (GP4) and note for SW version(SV25.2_R01 and above) SNDOC2398241 Create new topic for 1.5T HD Quad Extremity Coil Pre-Amp Replacement Procedure SN DOC 2301878 Update PM Schedule Overview FRU update for SCIM, SV25.3 software, PP filter SPT topics update Add content for Image Protocol Management Add content for Adjusting patient handling |
| 4 | March 10th, 2021 |
Add new FRU: ICN 5941000-9 Dell R640 (Gen7-DL) Add new FRU: New XFD-PS FRU 5549353-2 and New XFD-PS control board FRU 5837328.Add New XFD-PS Low level FRU Add New XFD-PS Low level FRU Add SV29.1/SV25.4 software installation. Add new content for ICW upgrade mode. |
| 5 | July 8th, 2021 |
Add Calibrating HOEC. |