RevisionDateReason For Change
1Nov 16, 2018Initial Release
2Oct 21, 2019

Chapter: Installation=>Guided Mechniacal Install Flow=>2rd day:cable connection.

Correct the diagram of "Install Temperature & Humidity Sensor on GOC" run1321 to 1383.

Add "Installing T5820 GOC" procedure. For "Installation of Fiber Optic Cables" add T5820 configurtion procedure and this procedure is same with T5810.

Add T5820 cable map.

Add GOCAA cable connection.

Chapter:Installation=>Install and Calibration Wizard for R/RD Series Magnet

In the calibration guided flow, figure 10 6th day, update "Sys Gain Cals" to "Sys Gain Cals (only for software SV25.2 and before system)" .

In the calibration guided flow, figure 10 6th day, click into the link of "SPT Full Test", in section 3 "Invoking SPT Full Test Mode", step 2, add (For SV25.2 and before system),.... (For SV25.3 and above system), click start SPT to begin, the following items "Setup check" "EC Check" "SNR/Sysgain""Stability" "CNT" will be proceed automatically.

Chapter:Adjust=>Cal=>System Level Procedures=>LVShim Procedures

For SN2270524. Add words: For the software SV25.2 and above System, The value of grad shim X/Y/Z will looks bigger than before because the resolution for Gradshim values changed (shift by 4 bits).

Chapter: Sys Diagrams=>System Block Diagram

Add ICN7 cabinet diagram

Chapter: Software=>Insite Remote Service Platform (RSvP) theory of operation

New create topic group

Chapter: Software=>System Host Software=>Loading host system software

Update "Guided install Site Configuration Options" list. In system configuration, add items "Hospital name; Suite ID; Host ID; Unique / System ID; Service ID", remove item "Enable HO Shim"; In Hardware configuration, add items "Scan range; Wireless PAC Frequency Channel", update "Scanner Channel configuration: 8 for Creator, 16 for Explorer"; Change "HEC Type/Cooling Cabinet Type" to "Cooling Cabinet Type"; Move "UPM Type" under "Table Configuration"; Change "BRM/BRM2" to "BRM,NOTE: BRM stands for BRMx coils"; Add below items: "Detachable Table: This is for motors table only, not applicable for SIGNA Creator and SIGNA Explorer." "Number of Tables: This is for motors table only, not applicable for SIGNA Creator and SIGNA Explorer." "Table Encoder Calibration Value: Not applicable for SIGNA Creator and SIGNA Explorer."" Encoder Calibration value 2: Not applicable for SIGNA Creator and SIGNA Explorer." "Altitude in meters (0 to 5000): Not applicable for SIGNA Creator and SIGNA Explorer."

update the pic "Sample of software revision to install" by removing advanced mode.

Chapter: Software=>System Host Software=>Loading USB host system software (Quick Start)

New create topic

Chapter: Theory=>RF Subsystem=>Receive Chain Theory

Update the theory diagram for ICN7

Chapter: Troubleshooting=>Diagnostics=>System Function=>Recon=>ICN Quick Health Check=>ICN Hardware Health for Dell ICN

Add R640 ("GEN7") in the senstance "The ICN Hardware Health Diagnostic is used on the Dell R620 ("Gen 5") and R630 ("Gen 6") ICNs only"

Chapter: Troubleshooting=>Diagnostics=>System Function=>Recon=>ICN Hardware Diagnostics=>ICN Hardware Intensive Diagnostics

Add R640 ("GEN7") in the senstance "The ICN Hardware Health Diagnostic is used on the Dell R620 ("Gen 5") and R630 ("Gen 6") ICNs only"

Chapter: Parts=>Illustrated parts=>PATIENT TRANSPORT LITE TABLE

For SN2302216. Update "Straps and Pads for Patient Table" FRU list

Chapter: Parts=>Illustrated Parts=>Low Hight Fixed Patient Table

Add "5458804-2" for Table Control Box

For SIMs feedback ID2594: update 5420619 description to "RF Cable FRU kit (RF Cable + Cable Clamp Assy)"

For SN2302216. Update "Straps and Pads for Patient Table" FRU list.

Update all the FRU list and structure.

Chapter: Parts=>Illustrated parts=>OC

Add T5820 PC FRU list.

Chapter: Parts=>Illustrated parts=>System Cabinet

Add GEN7 ICN in item13 and add table13. Add GP4 new FRU part 6237000-3

Chapter: Replacement=>Operator WorkSpace Replacement=>Dell T5820 computer replacement and Low Level FRU replacement

New create topic group

Chapter: Replacement=>System Cabinet=>ICN (Dell R640)

New topic group including "ICN Replacement", "ICN SSD Replacement(Dell R-640)" "ICN Power Supply Replacement(Dell R640)" "ICN On/Off procedure for Dell R640(Gen 7)"

Chapter: Replacement=>Low Height Fixed Table=>Table Control Box

Add a note "If Table Control Box 5458804 installed in site need to be replaced, please order 5458804/5458804-2 for replacement. If Table Control Box 5458804-2 installed in site need to be replaced, please order 5458804-2 for replacement."

Chapter: Replacement=>Coil Blower and Chiller Assemblies=>11kw Chiller Controller Replacement

For SN2194811. New create topic.

Chapter: Planned Maintenance

Update PM list.

3Sep 23th, 2020

SNDOC2312335 Update PM lists.

SNDOC2325945 Update Coil FRUs

SNDOC2302216 Patient handling accessories (pads straps) FRU introduction

SNDOC2326404 Update LHFT FRU list and structure

SNDOC2323031 Add 5376361 & 5376361-2 encoder box FRU

SNDOC2296770 Add 6237000-3 (GP4) and note for SW version(SV25.2_R01 and above)

SNDOC2398241 Create new topic for 1.5T HD Quad Extremity Coil Pre-Amp Replacement Procedure

SN DOC 2301878 Update PM Schedule Overview

FRU update for SCIM, SV25.3 software, PP filter

SPT topics update

Add content for Image Protocol Management

Add content for Adjusting patient handling

4March 10th, 2021

Add new FRU: ICN 5941000-9 Dell R640 (Gen7-DL)

Add new FRU: New XFD-PS FRU 5549353-2 and New XFD-PS control board FRU 5837328.Add New XFD-PS Low level FRU

Add New XFD-PS Low level FRU

Add SV29.1/SV25.4 software installation.

Add new content for ICW upgrade mode.

5July 8th, 2021

Add Calibrating HOEC.