- SIGNA™ Hero 3.0T Service Methods
- 5852800-8EN Revision 1.0
- 00000018WIA30D8E350GYZ
- id_20209541.3
- Aug 9, 2021 4:31:40 PM
Replacing the P2/P4 cable assembly
Removes the P2/P4 cable from the dockable patient table and installs a replacement cable assembly.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 2 | - | 8 hours | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 7.31 x 0.184 Self-Locking Cable Tie | As required | 46-208758P3 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| 3T P2/P4 Cable Track Assembly | 1 | 5807539-20 | - |