• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30B77C40GYZ
  • Topic ID: id_2019629 Version: 1.2
  • Date: Sep 19, 2019 12:05:14 PM

Replacing the dockable table ElectroMagnetic (EM) clutch and belt

Removes the ElectroMagnetic (EM) clutch and belt from the dockable table motor and EM clutch assembly, and installs a replacement clutch and belt.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-90 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Stainless Steel Ruler146-198452P2-
Tension Meter Tool15834577-
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Consumables
ItemQuantityPart numberManufacturer
Loctite #243As needed5415261-3-
Replacement parts
ItemQuantityPart numberManufacturer
Dockable Table Electromagnetic Clutch FRU Kit15819401-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.