Lowering the dockable table cradle

Lowers the cradle onto the dockable tabletop after service.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
2-15 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-

Procedure

  1. Remove the two support pads and lower the cradle onto the tabletop.
    Figure 1. Support pads
  2. Align the two holes in the cradle with the holes in the rigibelt mounting block.
  3. Install the two washers and two screws securing the cradle to the tabletop.
    Figure 2. Cradle hardware alignment
    1Screw
    2Washer
    3Rigibelt mounting block