• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA302B0450GYZ
  • Topic ID: id_2021636 Version: 1.2
  • Date: Jan 16, 2020 4:43:22 PM

Replacing a table caster

Removes one of the caster assemblies from the table base and installs a replacement caster.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-20 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
150 mm Plastic Dial Caliper with Metric Readout15916010-
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Scissor Jack15437045-
PH Alignment and Magnet Leveling Kit15897979-
Consumables
ItemQuantityPart numberManufacturer
Permatex (1 oz tube) Antiseize LubricantAs required2119594-
Loctite #243As required5415261-3-
Replacement parts
ItemQuantityPart numberManufacturer
White Caster Wheel with Ball Bearing Swivel15338208-5-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.