Checking the peripheral pulse probe
Make sure that the fiber-optic peripheral pulse probe functions correctly.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 5 minutes | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Procedure
Results
This illustration shows a peripheral pulse line graph when the probe is connected and working properly.

This illustration shows a peripheral pulse line graph when the probe is not connected.