Installs the Posterior Array (PA) coil into the patient cradle.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 45 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
Procedure
- Install the PA coil into the cradle.
- Install six screws securing the coil.
Note: Install the middle two screws prior to the four corner screws.
Figure 1. PA coil screws
- Connect cable connectors C1, and C2, and C3 to the PA coil.
Figure 2. PA coil connectors