Replacing the hydraulic rotary pump

Replaces the hydraulic rotary pump of the dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-90 to 105 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
11/16 inch Open-End Wrench2--
7/8 inch Open-End Wrench1--
Needle-Nose Pliers1--
Slotted Screwdriver1--
Consumables
ItemQuantityPart numberManufacturer
Loctite #24250 cc bottle--
Absorbent TowelsAs needed--
Hose Clamp146-317179P1 -
Replacement parts
ItemQuantityPart NumberManufacturer
Patient Table Hydraulic Rotary Pump Assembly15341568-
Required conditions
Table bellows must be raised.
Lower base side covers must be removed.
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Undock the patient table.
  2. Remove the patient table from the magnet room.