• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30CE5C40GYZ
  • Topic ID: id_2019356 Version: 2.3
  • Date: Apr 23, 2020 8:51:50 PM

Removing the dockable table bellows cover

Removes the bellows cover from the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-20 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Open the bellows cover zipper.
  2. Disconnect each rod, one at a time.
    Figure 1. Bellows cover rods
  3. Remove the screws securing the bellows cover to the table base covers.
    Figure 2. Bellows lower screws
  4. Remove the screws securing the bellows cover to the table side covers.
    Figure 3. Bellows upper screws
  5. Remove the cover from the table.