- SIGNA™ Hero 3.0T Service Methods
- 5852800-8EN Revision 1.0
- 00000018WIA30420130GYZ
- id_123732812.4
- Dec 28, 2021 6:19:07 PM
Replacing the hydraulic cylinder
Removes the hydraulic cylinder from a dockable patient table and installs a replacement hydraulic cylinder.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 90 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Flat-Blade Screwdriver, Small | 1 | - | - |
| Flat-Blade Screwdriver, Medium | 1 | - | - |
| Needle-Nose Pliers | 1 | - | - |
| 3/4 inch Open-End Wrench | 2 | - | - |
| Allen Key Set - Metric | 1 | 5115275 | - |
| Consumables | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Self-locking Cable Tie | As required | - | - |
| Absorbent Towels | As required | - | - |
| Bostik (4 oz can) Antiseize Lubricant or Permatex (1 oz tube) Antiseize Lubricant | 1 | 46-294151P8 or 2119594 | - |
| Hose Clamp | 1 | 46-317179P1 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Hydraulic Cylinder Assembly | 1 | 5142785-205 | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Procedure
- Undock the patient table.
- Remove the patient table from the magnet room.