• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30420130GYZ
  • Topic ID: id_12373281 Version: 2.4
  • Date: Dec 28, 2021 6:19:07 PM

Replacing the hydraulic cylinder

Removes the hydraulic cylinder from a dockable patient table and installs a replacement hydraulic cylinder.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-90 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Flat-Blade Screwdriver, Small1--
Flat-Blade Screwdriver, Medium1--
Needle-Nose Pliers1--
3/4 inch Open-End Wrench2--
Allen Key Set - Metric15115275-
Consumables
ItemQuantityPart numberManufacturer
Self-locking Cable TieAs required--
Absorbent TowelsAs required--
Bostik (4 oz can) Antiseize Lubricant or Permatex (1 oz tube) Antiseize Lubricant 146-294151P8 or 2119594-
Hose Clamp146-317179P1 -
Replacement parts
ItemQuantityPart numberManufacturer
Hydraulic Cylinder Assembly15142785-205-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Undock the patient table.
  2. Remove the patient table from the magnet room.