• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30B7E030GYZ
  • Topic ID: id_12373721 Version: 2.5
  • Date: Jan 26, 2022 4:01:13 PM

Replacing the hydraulic filter

Removal and installation instructions for replacing the hydraulic filter in the dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-40 minutes -
Tools and test equipment
ItemQuantityPart numberManufacturer
3/4 inch Open-End Wrench2--
Ear Syringe1--
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Consumables
ItemQuantityPart numberManufacturer
Hydraulic Oil1-
TowelsAs needed--
Loctite® 5691 bottle (50 ml)46-230614P1-
Replacement parts
ItemQuantityPart numberManufacturer
Hydraulic Oil Filter And Gasket1-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

Undock the patient table and remove it from the magnet room.