• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30BF5C40GYZ
  • Topic ID: id_2019375 Version: 2.4
  • Date: Apr 23, 2020 8:51:51 PM

Installing the dockable table bore-end cradle cover

Installs the bore-end cover onto the dockable table cradle.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Position the bore-end cradle cover at the bore end of the cradle.
    Figure 1. Bore-end cradle cover
  2. Install the two screws securing the microphone to the cover.
    Figure 2. Bore-end cradle cover component screws
    1P-port screw
    2Microphone screw
  3. Make sure the port covers move with no interference, then install the four screws securing each P-port to the cover.
    Note: Lift the bore-end cover and move any interfering cables, then lower the cover and retest the port covers if required.
  4. Install the cover onto the cradle.
  5. Install the four screws securing the cover to the cradle.
    Figure 3. Bore-end cradle cover screws
  6. Push the cradle back to the home position on the tabletop.
  7. Dock the table.