• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30F67E40GYZ
  • Topic ID: id_2020306 Version: 2.1
  • Date: Dec 13, 2019 10:34:42 AM

Replacing the cradle flipper assembly

Replaces the cradle flipper assembly in a dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-60 minutes -
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
Consumables
ItemQuantityPart numberManufacturer
Loctite #243As required5415261-3-
Replacement parts
ItemQuantityPart numberManufacturer
Flipper Assembly15796733-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Undock the patient table.
  2. Remove the patient table from the magnet room.