• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA309B1E40GYZ
  • Topic ID: id_2019681 Version: 2.0
  • Date: Oct 11, 2019 11:04:36 AM

Replacing the dockable table string encoder

Removes the string encoder from the dockable table and installs a replacement encoder assembly.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-4 hours and 25 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
Replacement parts
ItemQuantityPart numberManufacturer
Dockable Table String Encoder FRU15796731-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.