Installing the dockable table bellows cover

Installs the bellows cover onto the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

Procedure

  1. Install the bellows cover onto the table.
  2. Connect each rod, one at a time.
    Figure 1. Bellows cover rods
  3. Raise the bellows and secure the bellows cover to the table side covers with M6 x 20 Button Head Collar Screws.
    Figure 2. Bellows upper screws
  4. Pull the bellows cover zipper closed.
  5. Secure the bellows cover to the table base side covers with M6 x 20 Button Head Collar Screws.
    Figure 3. Bellows lower screws