Removes the primary latch assembly from the dockable table and installs a replacement primary latch assembly.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
2 hours and 20 minutes
5 minutes
Tools and test equipment
Item
Quantity
Part number
Manufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set
1
5113258
-
Stainless Steel Ruler
1
46-198452P2
Consumables
Item
Quantity
Part number
Manufacturer
Loctite #243
As required
5415261-3
-
Replacement parts
Item
Quantity
Part number
Manufacturer
Primary Latch Assembly
1
5796734
-
Required conditions
Table is undocked and away from the magnet.
Safety
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
Comply with all site-specific training and workplace safety requirements.
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.
CAUTION
Potential pinch hazard
Fingers placed in the path of moving components when checking the adjustment or operating the mechanism can be pinched.
Keep fingers clear of the moving components when checking the adjustment or operating the mechanism
Procedure
Undock the table and move it away from the magnet. If possible, move it out of the scan room.