Removes the table dock electrical motor and installs a replacement motor.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 40 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Consumables |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Permatex (1 oz tube) Antiseize Lubricant | As required | 2119594 | - |
| Loctite #243 | As required | 5415261-3 | - |
| Replacement parts |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Dock Motor | 1 | 5173426 | - |
| Required conditions |
|---|
| The dock cover is removed. |
| Warning |
|---|
 | Ferrous material hazard The table dock components are ferrous and can become dangerous projectiles within the magnetic field. The dock must be removed from the magnet room prior to removing any parts from the dock. |