• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30B02840GYZ
  • Topic ID: id_2014696 Version: 4.0
  • Date: Feb 26, 2021 3:51:10 AM

Installing the Integrated Control Engine (ICE) fan assembly

Install the Integrated Control Engine (ICE) fan assembly in the System Cabinet.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 to 10 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set1 Kit5113258-
Nonmagnetic Titanium Service Tool Kit, Large Set5112581-
Replacement parts
ItemQuantityPart numberManufacturer
pICE fan2 5485930-

Procedure

Do the following steps to install a fan in the back of the ICE.
  1. Connect the fan cable to the power cable in the hole beside the fan slot.
    Note: Be careful that the power cable does not fall into the hole.
    Figure 1. Hole for power cable
  2. Tighten four screws to secure the fan to the ICE.
Figure 2. Installing the fan
1Connect the fan cable to the power cable
2Tighten four screws