Install the Integrated Control Engine (ICE) fan assembly in the System Cabinet.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 5 to 10 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 Kit | 5113258 | - |
| Nonmagnetic Titanium Service Tool Kit, Large Set | 5112581 | - |
| Replacement parts |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
|
pICE fan | 2 |
5485930 | - |
Procedure
Do the following steps to install a fan in the back of the ICE.- Connect the fan cable to the power cable in the hole beside the fan slot.
Note: Be careful that the power cable does not fall into the hole.
Figure 1. Hole for power cable
- Tighten four screws to secure the fan to the ICE.
Figure 2. Installing the fan
| 1 | Connect the fan cable to the power cable |
| 2 | Tighten four screws |