Leak Sensor 2 (Front)

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable10 minutes15 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Standard Tool1---
Table 3. Replacement parts
ItemQuantityEffectivityPart numberManufacturer
Leak Sensor 2 (Front) (Refer to Illustrated Parts)1---
Table 4. Safety
Table 5. Required conditions
ConditionReferenceEffectivity

System Power must be turned OFF. Refer to System Cabinet PDU Main Breaker LOTO Procedure.

--

Procedure

  1. Remove L Upper Front Cover. Refer to SC Cover Removal
  2. Disconnect the two connectors of leak sensor 2 (Front).
  3. Remove the leak sensor 2 (Front) by loosening 6 screws.
    Figure 1. Leak Sensor (Front)
  4. Installation is reverse order of removal.