Starting the coil datapath diagnostic

Diagnose the datapath connection for the coil by testing the connections between the P-coil connectors and the Digitized Per Pin Remote Receiver (DPP RRx) modules, between the PA port E connectors and the DPP RRx modules, and between the optional Multi-Nuclear Spectroscopy (MNS) frequency converter and the DPP RRx module if installed.

About this task

Complete the coil datapath diagnostic procedure to make sure that the signal paths between the inputs are functioning correctly.
Figure 1. Dockable table coil connector and port locations
Connection point locations
CalloutConnection points
1P1 (16 channel)
2P2 (32 channel)
3PA Port E (32 channel); connection points C1, C2, C3
4P4 (16 channel)
5DPP RRx1
6DPP RRx2
7DPP RRx3
Figure 2. Signal paths between inputs