Removes the coupler from the rigibelt drive shaft.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 5 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Required conditions |
|---|
| The motor and ElectroMagnetic (EM) clutch assembly is removed from the table. |
Procedure
- Loosen the screw securing the motor coupler to the EM clutch drive shaft.
Figure 1. Motor coupler screw
- Remove the motor coupler from the EM clutch shaft.