Adjusting the patient table belt tension

Measures and adjusts belt tension for a dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Stainless Steel Ruler146-198452P2-
Tension Meter Tool15834577-
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-

Procedure

  1. Put the small o-ring on the tension meter at zero.
    Figure 1. Small o-ring at zero
  2. Put a ruler on the motor and ElectroMagnetic (EM) clutch assembly base close to the center of the belt span.
    Figure 2. Belt tension setup
    1Belt
    2Tension meter tool
    3Ruler
  3. Put the tension meter on the belt squarely at the center of the belt span with the ruler visible.
  4. Align your point of view such that you are looking at the belt directly on its edge with the scale of the ruler behind the belt.
  5. Without changing your point of view, press the tension tool plunger straight down onto the belt until the belt has deflected 3 mm as measured on the ruler.
    Figure 3. Belt tension measurement
    1Deflected measurement
    2Undeflected measurment
    3Tension meter tool plunger
    4Tension meter tool
    5Ruler
    6Deflected belt
  6. Release the tension tool and read the force applied from the bottom of the small o-ring on the deflection force scale. Compare the force measurement to the table below.
    Figure 4. Belt tension reading o-ring
    Table 1. Belt tension
    Belt typeForce
    New1.0-1.1 kilograms
    Used0.8-1.0 kilograms
  7. If the belt tension is outside of minimum or maximum values, tighten or loosen the belt tension screw as follows:
    1. Slightly loosen the four screws securing the motor to the motor housing.
      Figure 5. Belt tension adjustment screws
      1Motor securing screws
      2Belt tension adjustment screw
    2. Turn the tension adjustment screw to adjust the belt tension. Turn clockwise to tighten the belt or counter-clockwise to loosen the belt.
    3. Tighten the four screws securing the motor to the motor housing.
  8. Repeat Step 1 through Step 7 until the deflection force is within tolerance.