• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30A80030GYZ
  • Topic ID: id_13105807 Version: 1.3
  • Date: Jul 5, 2019 11:49:11 PM

Gradient Filter

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable30 minutes15 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Non Magnetic Tool Set1---
Digital volt meter (DVM) with alligator clip leads1---
Table 3. Replacement parts
ItemQuantityEffectivityPart numberManufacturer
Gradient Filter (Refer to Illustrated Parts)1---
Table 4. Safety
Table 5. Required conditions
ConditionReferenceEffectivity

PDU must be turned OFF. Refer to Lockout / Tagout for PDU in System Cabinet.

--

Procedure

  1. Remove SC Rear cover of magnet room side. Refer to SC Cover Removal
  2. Remove the terminal Cover for defective Gradient Filter by removing 2 screws from gradient filter.
  3. Check the voltage between Ground Stud and each terminal (+ or -). Verify that there is no voltage.
  4. Remove the cable from defective gradient filter while keeping the spacer on cable intact.
    Note:

    Use two wrenches as illustration to tighten the nut. Need to tighten with enough torque to avoid spike noise. But, do not over tighten not to break stud.

  5. Cut tie-wraps at defective Gradient Filter.
    Figure 1. Remove the all cables
  6. Remove the 24 screws from frame of gradient filter.
  7. Remove gradient filter from Penetration Panel.
    Figure 2. Remove Gradient Filter
  8. Restore the Gradient Filter by tightening 24 screws.
    Note:

    Make sure to tighten each screw.

  9. Connect Gradient Cables. Make sure screw is fixed, nut is tight, and cable is not detached from terminal.
    Note:

    Use two wrenches to tighten the nut.

  10. Restore terminal Cover and restore Cabinet cover.