• Topic ID: id_12374577
  • Version: 3.0
  • Date: Oct 10, 2019 2:36:42 PM

Replacing the reciprocating pump assembly

Removes the reciprocating pump assembly from the dockable patient table and installs a replacement reciprocating pump assembly.

Prerequisites

Personnel requirements
Required persons Preliminary requirements Procedure Finalization
1 - 60 minutes -
Tools and test equipment
Item Quantity Part number Manufacturer
3/4 inch Open-End Wrench 1 - -
Needle-Nose Pliers 1 - -
Consumables
Item Quantity Part number Manufacturer
Absorbent Towels As needed - -

There are two types of Reciprocating Pump Assemblies depending on the weight rating of the table. The GEM flat table requires the High Pressure Reciprocating Pump, while all curved tables use the Standard Reciprocating Pump.

Replacement parts
Item Quantity Part number Manufacturer
Reciprocating Pump Assembly 1 See FRU manual -
High Pressure Reciprocating Pump Assembly 1 See FRU manual -
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.