- id_12373738
- Version: 1.3
- Date: Jul 5, 2019 6:08:29 PM
Hydraulic Fluid Draining and Rotary Pump Replacement Videos
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | 5 minutes | 120 mins. for Pump replacement; 90 mins. for Hydraulic fluid replacement; 15 mins. for Air bubble extraction | 5 minutes |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Hydraulic Fluid Container. Maximum height 3.5 inches (90 mm), must hold 64 oz. (2 liters). Video example is 2 in. high x 6 in. wide x 12 in. long. | 1 | - | - | - |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Hydraulic Fluid | 2 qts | - |
See FRU manual |
- |
| Absorbent Towels | As Needed | - |
N/A |
- |
| Wire Ties | As Needed | - |
N/A |
- |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Rotary Pump | 1 | - |
See FRU manual |
- |
|
Overview
Visit the MR Service Documentation Website to view three different video clips showing the process to:
-
Replace the rotary pump
-
Drain the table hydraulic fluid
-
Remove air bubbles
These videos are for reference only and are not intended to replace the existing service manual procedures. Refer to Patient Table Hydraulic Rotary Pump Replacement and Hydraulic Fluid Level Check and Fill for detailed instructions.
Procedure
- Replace the rotary pump. Refer to the Replace Rotary Pump video on the MR Service Documentation Website.
- Replace the hydraulic fluid only. Refer to the Replace
Hydraulic Fluid Only video on the MR Service Documentation
Website.note:
Three points to be aware of during the video:
-
To unclip the hose clamp, twist the clip ends in opposite directions
-
Note the point in the video when the hose end is covered to prevent hydraulic fluid from draining
-
The fluid height in the reservoir will change depending on the height of the table. For the final fill, the table is raised to full height—ensure the reservoir is only half-full
-
- Extract air bubbles from the hydraulic lines. Refer to the Extract Air Bubbles video on the MR Service Documentation Website.
Finalization
No finalization steps.