• Topic ID: id_12373281
  • Version: 2.0
  • Date: Oct 10, 2019 11:40:29 AM

Replacing the hydraulic cylinder

Removes the hydraulic cylinder from a dockable patient table and installs a replacement hydraulic cylinder.

Prerequisites

Personnel requirements
Required persons Preliminary requirements Procedure Finalization
1 - 90 minutes -
Tools and test equipment
Item Quantity Part number Manufacturer
Flat-Blade Screwdriver, Small 1 - -
Flat-Blade Screwdriver, Medium 1 - -
Needle-Nose Pliers 1 - -
3/4 inch Open-End Wrench 2 - -
Consumables
Item Quantity Part number Manufacturer
Self-locking Cable Tie As required - -
Absorbent Towels As required - -
Replacement parts
Item Quantity Part number Manufacturer
Hydraulic Cylinder Assembly 1 5142785-205 -
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.