Doing a compressor pressure check
Do a check of the cryocooler compressor supply pressure.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 5 minutes | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
Procedure
- Make sure the cryocooler system is at equilibrium (coldhead at approximately 4.2 K from Magnet Monitor 3).note: Do not do a check of the operating (dynamic high-side) pressure if the cryocooler system is not at equilibrium (coldhead is not at approximately 4.2 K). It will not be accurate.
- Do a check of the operating (dynamic high-side) pressure of the cryocooler system and compare the pressure to the specifications below.
Figure 1. Gauge for RDK-408A3 and RDK-408A2
note: The supply pressure gauge displays static pressure when the compressor is off and high-side dynamic pressure when the compressor is on. The table below shows the applicable compressor operating (dynamic high-side) pressure range for the compressor and coldhead combination.Table 1 Acceptable compressor operating pressure range Compressor model Compressor operating (dynamic high-side) pressure, MPa (approximate) F50H and F50L 1.90 to 2.20 CSA-71A and CSW-71D 2.00 to 2.20 If the dynamic high-side pressure is within the acceptable range, the task is complete. - If the dynamic high-side pressure is outside the acceptable range:
- Turn the cryocooler system off.
- Let the system stabilize.
- Do a check of the static pressure and compare the pressure to the specifications below.
Table 2 Compressor pressure specifications Coldhead model Compressor static gas pressure at 20°C (68°F), MPa (approximate) RRDK-408 4K two-stage coldhead 1.60 to 1.65 RDK-408, RDK-408A2, RDK-408A3 4K coldhead 1.60 to 1.65 RDK-408T, RDK-400 1.45 to 1.50 - If both the dynamic and static pressures are within the acceptable range, the task is complete.
- If both the dynamic and static pressures are out of range, continue on with the next steps for Adjusting the static pressure on the compressor.