- id_12374015
- Version: 1.7
- Date: Mar 3, 2020 12:42:30 PM
Replacing table electrical docking mechanism components
Removes table components related to electrical docking and installs replacement components.
Prerequisites
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 60 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Replacement parts | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| Link Bar Connector Pedal | 1 | 5174518 | - |
| Link Rod Connector | 2 | 5769330 | - |
| Shaft Rocker | 2 | 5147700-2 (GEM) 5147700 (Non-GEM) |
- |
| Rocker Clamp | 2 | 5392170 (GEM) 5147701 (Non-GEM) |
- |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must: