• Topic ID: id_12374633
  • Version: 1.23
  • Date: Oct 22, 2019 12:14:10 PM

Replacing the hydraulic rotary pump

Replaces the hydraulic rotary pump of the dockable patient table.

Prerequisites

Personnel requirements
Required persons Preliminary requirements Procedure Finalization
1 - 90 to 105 minutes -
Tools and test equipment
Item Quantity Part number Manufacturer
Nonmagnetic Titanium Service Tool Kit, Large Set 1 5112581 -
11/16 inch Open-End Wrench 2 - -
7/8 inch Open-End Wrench 1 - -
Needle-Nose Pliers 1 - -
Slotted Screwdriver 1 - -
MCRv Tool Kit 3.0T 1 5182417-2 -
MCRv Tool Kit 1.5T 1 5182417 -
Consumables
Item Quantity Part number Manufacturer
Loctite #242 50 cc bottle - -
Absorbent Towels As needed - -
Replacement parts
Item Quantity Part Number Manufacturer
Patient Table Hydraulic Rotary Pump Assembly 1 5341568 -
Patient Table Hydraulic Rotary Pump Female Coupler 1 5336201 -
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

  • warning
  • Personal injury possibility
  • When doing maintenance on the table in its up position, severe personal injury can occur if the table is accidentally lowered during maintenance.
  • Make sure the lock safety bar is safely in place.