Installing the dockable table interface cover

Installs the dockable table interface cover onto the magnet front cover.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
Front endbell is installed.
Left- and right-hand lower access panels are removed.

Procedure

  1. If docked, raise the patient table to the up-limit position.
  2. Install the dockable table interface cover onto the magnet front cover.
    Figure 1. Dockable table interface cover
  3. Install the four M10 x 25 Socket Head Cap Screws securing the dockable table interface cover to the magnet front cover.
    Figure 2. Dockable table interface cover screws