• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30D59350GYZ
  • Topic ID: id_2020647 Version: 1.8
  • Date: Apr 23, 2020 8:53:09 PM

Installing the dock

Installs the magnet-side dock.

Prerequisites

Procedure

  1. With two field engineers, one holding each side of the dock, enter the magnet room and place the dock on the floor where the foot of the patient table would normally be located.
    Note: Route the dock cables under the magnet to the service side of the magnet while moving the dock.
  2. When the dock has been moved to the normal location of the patient table foot, one person can apply constant downward pressure to hold the dock in place while the other person carefully installs the cables in the front of the magnet.
  3. With two people, one holding each side of the dock, slowly and deliberately slide the dock towards the magnet, keeping downward pressure on the top of the dock assembly.
  4. Use two people to position the dock immediately to the side of the anchor stud.
  5. Use two people to lift the front of the dock JUST ENOUGH to clear the top of the stud in the floor. Avoid scraping the dock assembly or the base of the magnet.
  6. Install four sets of spherical washers, one set from each of the mounting studs.
    Figure 1. Dock mounting hardware
    1Spherical washer sets
    2Lock washer
    3Nut
  7. Install the dock to the dock mounting bracket.
  8. Install four M10 Stainless Steel Hex Nuts (3000-M10C-33), four M10 Bronze Lock Washers (7700820-210), and four spherical washer sets securing the dock to the dock mounting bracket.
  9. If the site has followed the Pre-Installation Manual and the anchor is a two-part assembly, install the anchor stud (bolt) to the female sleeve anchor, using hand-force to turn the stud. Install the nut, washers, and clamp bracket on the anchor stud.