• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WHA30A4A4GYZ
  • Topic ID: id_2001797 Version: 1.17
  • Date: Jul 5, 2019 7:48:58 AM

Replacing the DVD drive on the host computer

Removal and installation instructions for replacing the DVD drive on the host computer.

Prerequisites

The following tables represent a comprehensive roll-up of the tools, test equipment, consumables, and replacement parts you need to complete all possible tasks in this procedure.

Tools and test equipment
ItemQuantityPart numberManufacturer
Screwdriver1--
ESD Mat1--
ESD Strap1--
4 mm Hex Key Wrench1--
Brass Master Padlock with Brass Shackle246-194427P320-
Red LOTO Personal Lock Wrap12393068-
Multi-Locking Device (if multiple service personnel are involved)146-194427P313-
Red Warning LOTO Tag146-194427P322-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.