• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA308D4840GYZ
  • Topic ID: id_2014923 Version: 4.0
  • Date: Dec 6, 2019 3:19:21 AM

Replacing the CMOS battery - Dell T5820

Removal and installation instructions for replacing the CMOS battery in the host computer.

Prerequisites

The following tables represent a comprehensive roll-up of the tools, test equipment, consumables, and replacement parts you need to complete all possible tasks in this procedure.
Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-60 minutes10 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Screwdriver1--
Nonmagnetic Titanium Service Tool Kit, Small Set (can be used on magnets with magnetic fields > 1.5T)15113258-
ESD Mat1--
ESD Strap1--
DQA III Phantom12321556-
DQA Phantom Positioner15554497-
Consumables
ItemQuantityPart numberManufacturer
GE Equipment Maintenance Sticker15661793-
Replacement parts
ItemQuantityPart numberManufacturer
Dust Filter1 (if needed)5192248-5-
Adhesive-Backed Velcro, 2 x 4 inches2 (if needed)46-317475P2-
Host PC CMOS Battery, 3V, CR203215129534-3-
Required conditions

Use an antistatic wrist strap and mat while performing this procedure to prevent electrostatic discharge (ESD).

Access to both the front and rear of the operator console is required. Move the console away from walls and remove obstacles in the room.

Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.