• SIGNA MR355 / SIGNA MR360
  • Service Manual
  • 5856356-3EN Revision 5.0
  • Basic Service Documentation. Copyright General Electric Company.
  • Object ID: 00000018WIA30124840GYZ
  • Topic ID: id_2014880 Version: 3.0
  • Date: Dec 9, 2019 1:20:44 AM

Replacing the host computer - Dell T5820

Removal and installation instructions to replace the host computer when upgrading the system.

Prerequisites

The following tables represent a comprehensive roll-up of the tools, test equipment, consumables, and replacement parts you need to complete all possible tasks in this procedure.
Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-95 minutes2 hours
Tools and test equipment
ItemQuantityPart numberManufacturer
Screwdriver1--
Disk Management Tool1--
ESD Strap1--
Brass Master Padlock with Brass Shackle246-194427P320-
Red LOTO Personal Lock Wrap12393068-
Multi-Locking Device (if multiple service personnel are involved)146-194427P313-
Red Warning LOTO Tag146-194427P322-
DQA III Phantom12321556-
DQA Phantom Positioner15554497-
Consumables
ItemQuantityPart numberManufacturer
GE Equipment Maintenance Sticker15661793-
Replacement parts
ItemQuantityPart numberManufacturer
Dell T5820 MR Global Host W-2123 64GB Nvidia SSD18780011-02-
GOC Dust Filter for T58202 (if needed)5793940-
Adhesive-Backed Velcro, 2 x 4 inches2 (if needed)46-317475P2-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.