Removing the bellows

Detaches bellows cover from the side covers of the dockable patient table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-20 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-

Procedure

  1. Remove the screws securing the bellows cover to the table side covers.
    Figure 1. Bellows upper screws
  2. Lower the bellows cover.
  3. Remove the screws securing the bellows cover to the table base covers.
    Figure 2. Bellows lower screws