Removes the Posterior Array (PA) coil from the patient table cradle.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 35 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Required conditions |
|---|
| Cradle hardtop cover is removed. |
| LOTO - ISC is applied or the dockable table is undocked. |
Procedure
- Make sure the table is undocked prior to disconnecting the cable connectors.
| Notice |
|---|
 | Equipment damage MEMS circuitry in the posterior array coil may be damaged. Connect the C3 connector to the PA coil last. Disclaimer: Newly installed or updated systems may or may not have the MEMS power supply. |
Disconnect cable connectors C3, C2, and C1.
Figure 1. PA coil connectors
- Remove six screws securing the PA coil.
Figure 2. PA coil screws
- Remove the PA coil.