• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WHA301B62GYZ
  • Topic ID: id_2001423 Version: 1.28
  • Date: Jul 13, 2021 3:56:30 PM

Removing the Posterior Array (PA) coil

Removes the Posterior Array (PA) coil from the patient table cradle.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-35 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
Cradle hardtop cover is removed.
LOTO - ISC is applied or the dockable table is undocked.

Procedure

  1. Make sure the table is undocked prior to disconnecting the cable connectors.
  2. Disconnect cable connectors C3, C2, and C1.
    Figure 1. PA coil connectors
    1C1
    2C2
    3C3
  3. Remove six screws securing the PA coil.
    Figure 2. PA coil screws
    1Screw
    2PA coil
  4. Remove the PA coil.