• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30C0B350GYZ
  • Topic ID: id_2020840 Version: 1.13
  • Date: Apr 23, 2020 8:52:44 PM

Removing the dockable table cradle

Removes the cradle from the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
2-30 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
The P1/P3/PE and P2/P4 cables are removed from the cradle.

Procedure

  1. Make sure all cables from the table are disconnected from and not routed through the bottom of the cradle.
  2. Remove the following from the cradle:
    • The rear flap and supports
      Figure 1. Rear flap and supports
    • Four P3 and P4 balun holders
      Figure 2. Balun holders
    • The Electro-CardioGram (ECG) connector bracket
      Figure 3. ECG connector bracket
    • The secondary latch interface
      Figure 4. Secondary latch interface
    • 24 cradle vertical wheels
      Figure 5. Vertical wheels
    • Eight cradle side wheels
      Figure 6. Side wheels
    • Two hardtop cover stop screws
      Figure 7. Hardtop stop screws
    • Two cover plates
      Figure 8. Cover plates
    • The primary latch interface
      Figure 9. Primary latch interface
  3. Remove the cradle from the table.