• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA309FCE20GYZ
  • Topic ID: id_13107454 Version: 2.0
  • Date: Feb 22, 2021 1:47:19 AM

3.0T 8-Channel Wrist Coil Setup for MCQA Test

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable15 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
3.0T HD 8-Channel Wrist Array Legacy Phantom1Legacy phantom set

5160986-5

-
Cubical Unified Phantom1Unified phantom set

5342681

-
Wrist Array Baseplate with Baseplate Riser1-

5160986-12

-
Table 3. Required conditions
ConditionReferenceEffectivity

For Discovery MR750, the following coil configuration names must be installed to run this tool: GE_HDx Wrist.

--

About this task

Follow this process to prepare for the MCQA test using the 3.0T 8-Channel Wrist Array by Invivo, catalog M3340CA.

  • (For curved table) 3.0T HD 8-Channel Wrist Array by Invivo, catalog number M3340CA

  • (For flat table) 3.0T HD 8-Channel Wrist Array by Invivo, catalog number M7000GW

Note:

Coils do not ship with phantoms. Phantoms come in a unified phantom set with the MR system.

Topic ID: id_SL3233507-1104209

Legacy Phantom Procedure

Procedure

  1. Note:

    If installing the coil for the first time on a system, refer to Auto Coil Installation.

    Place the baseplate on the table.
    Figure 1. Baseplate on Table
  2. Attach the baseplate riser to the baseplate. Align the knobs and lock by pulling back.
    Figure 2. Baseplate Riser on Baseplate
    Figure 3. Baseplate and Riser in Locked Position
  3. Attach the 3.0T HD 8-channel wrist array to the cradle. Use the knobs to align and lock the coil. Note the serial number of the coil.
    Figure 4. Placement of Coil on Baseplate Assembly
  4. Slide the locked coil to the center position and mechanically locate and align it. Line it up to the baseplate alignment marker.
    Figure 5. Locked Coil Alignment
  5. Place the phantom in the coil.
  6. Lock the coil using the lock on inferior top lock.
  7. Establish an axial landmark on the line on top of the coil. Advance to scan.
    Figure 6. Landmark Reference Marks
  8. Proceed to the Multi-Coil Quality Assurance Tool procedure.
Topic ID: id_SL3233510-1104209

Unified Phantom Procedure

Procedure

  1. Place the baseplate of the wrist coil on the patient table.
    Figure 7. Baseplate of Wrist Coil on Patient Table
  2. Attach the base plate riser to the base plate.
    Figure 8. Baseplate Riser Attached to Baseplate
    Figure 9. Baseplate and Riser in Locked Position
  3. Attach the 3.0T HD 8-channel wrist array to the baseplate assembly using the knobs. Slide the locked coil to center and align properly.
    Figure 10. Wrist Array Attached to Baseplate
  4. Place the cubical unified phantom in the coil by lifting the upper portion of the coil.
    Figure 11. Cubical Unified Phantom Positioned in Coil
  5. Close the coil and lock it.
    Figure 12. Coil Closed and Locked
  6. Connect the coil to port A of the LPCA. Landmark the coil at the marks and advance to scan.
    Figure 13. Lock and Landmark the Coil
  7. Proceed to the Multi-Coil Quality Assurance Tool procedure.
Topic ID: id_14930324

Finalization

Finalization

No finalization steps.