• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA30E6CE20GYZ
  • Topic ID: id_13107250 Version: 3.0
  • Date: Aug 29, 2019 1:45:54 AM

3.0T 8-channel high res brain array coil troubleshooting

Topic ID: id_SL2763915-584045

Personnel Requirements

Required Persons Procedure
130 minutes
Topic ID: id_SL2763916-584045

Overview

The following tips can be used to troubleshoot common problems with the 3.0T HD 8 Channel High Resolution Brain Array by Invivo (M3335LL).

Topic ID: id_SL2763917-584045

Tools and Test Equipment

Digital Volt Meter

Topic ID: id_SL2763918-584045

Required Conditions

The following coil configuration names must be installed to run this tool with 15.x and earlier:

  • HDx 8HRBrain

  • HDx 8HRBrainServ

Topic ID: id_SL2763921-584045

Procedure

Topic ID: id_SL2763922-584045

Receiving No Signal

Problem:

Unable to pre-scan or scanning, but receiving no signal.

Possible Solution:

  • Verify the green light above Port A is illuminated. This indicates the coil is properly plugged into the system.

  • Verify the system correctly detected the coil by checking in Currently Connected in the GUI shown in the example in Figure 1.

Figure 1. Currently Connected Coils Window on Scan Screen
  • Verify that the landmark is correct and the cradle did not unlatch.

  • Verify that the scan locations and any Field of View (FOV) offsets are correct.

  • Perform a continuity check on the external cable described in PIN Diode Control Channels Check. The continuity check must be performed by a GE-authorized Service Engineer.

  • Verify that the coil is positioned with the cable exiting towards the bore and that the system and coil polarity match.

If still unable to receive a signal, try to scan (transmit and receive) with the body coil. For this test, be sure to remove the imaging coil from the magnet bore before scanning with the body coil.

  • If still not receiving a signal, the problem probably lies with the MR system.

  • If the scan completes successfully, there may be a problem with the Invivo coil. Contact GE for further assistance. If unable to scan with the substitute coil, there may be a system problem related to this particular coil type.

Topic ID: id_SL2763923-584045

Image Quality

Problem:

Poor IQ, shaded images, or MCQA fails.

Possible Solution:

  • Perform a continuity check on the external cable described in External Cable Wear. The continuity check must be performed by a GE authorized Service Engineer.

  • Verify that there are no loops in the cables.

  • Verify that there are no metal or ferromagnetic objects close to the coil, patient or magnet (i.e., safety pin, hair pin).

  • Verify that the coil is properly positioned.

  • Verify that the center frequency is within the frequency adjustment range for the system.

  • Verify that the R1, R2 and TG values from the pre-scan are within normally expected ranges.

If not already done:

  1. Perform a system Quality Assurance phantom test.

    1. If the values obtained do not fall within normal operating parameters, investigate further by performing a phantom scan with the body coil. For this test, be sure to remove the imaging coil from the magnet bore before scanning with the body coil. If still having the same problems, there is probably an MR system problem.

    2. If the body coil scan is satisfactory, acquire a scan using another coil of the exact same type (receive-only, phased array, or transmit/receive, whichever applies) and the same system coil selection. If the image quality is visibly improved, there may be a problem with the Invivo coil.

  2. Contact GE for further assistance. If the image quality still suffers, there may be a system problem related to imaging with this type of coil.

Topic ID: id_SL2763930-584045

Artifacts

Problem:

There is a black line or signal void on the image.

Possible Solution:

Verify that there is no metal present in the area being scanned in or on the patient.

Problem:

Some or all of the images appear shaded or exhibit uneven signal or banding.

Possible Solution:

  • Confirm that no metallic objects are located nearby, outside the FOV. This is especially important on images utilizing Fat Saturation.

  • If Fat Saturation is being used, verify that the Center Frequency Adjustment (CFA) fine adjustment was optimized.

Topic ID: id_SL2763931-584045

External Cable Wear

Problem:

System will not recognize the coil, or reports MC bias errors with the coil attached.

Possible Solution:

Perform the cable continuity check as follows:

Topic ID: id_SL2763932-584045

RF Signal Channels Check

The center pins of the System Connector receive coaxial connectors are extremely fragile. Do not attempt to make any measurements at these center pins.

  1. With the coil cable disconnected, use a multimeter to assure an open circuit condition exists between the designated pin and GND (shown in Figure 2) for each Signal Channel in .

    Figure 2. Coil Cable Connector
    Table 1.
    Signal Channel Coil Cable Connector Connection
    MC-1Center A7GND
    MC-2Center A6GND
    MC-3Center A5GND
    MC-4Center A4GND
    MC-5Center A3GND
    MC-6Center A2GND
    MC-7Pin 2GND
    MC-8Center A1GND
    GNDNOTE: All outer [A] shells and Pins 1 and 3 are common to GND.
  2. If one or more of the readings indicate a short, replace the cable. If all of the above readings indicate an open circuit condition, proceed to PIN Diode Control Channels Check.

Topic ID: id_SL584006-584045

PIN Diode Control Channels Check

  1. With the coil cable disconnected, use a multimeter to determine continuity between the designated pins for each Control Channel in Table 2. Verify that the control pins are not shorted to GND.

    Table 2. External Cable Expected Readings B
    Control Channel Coil Cable Connector System Connector
    MC bias 1Pin 17J1
    MC bias 2Pin 16J2
    MC bias 3Pin 15J3
    MC bias 4Pin 14J4
    MC bias 5Pin 4J5
    MC bias 6Pin 5J6
    MC bias 7Pin 6J7
    MC bias 8Pin 7J8
    GNDNOTE: All outer [A] shells and Pins 1 and 3 are common to GND.K1 - K8, M6
  2. If one or more of the readings are >3 ohms, replace the coil cable. If all readings are <3 ohms, proceed to PIN Diodes Check.

Topic ID: id_SL2763936-584045

PIN Diodes Check

  1. With the coil cable connected to the coil, use a multimeter to measure the diode drop across the coil connector pins as detailed in Table 3. Use Figure 3 for pin identification while performing this procedure.

    Table 3. Pin Diode Expected Readings
    System Connector Reading
    Positive Lead Negative Lead
    J1 GND[K1]1.7
    GND[K1]J1 Open
    J2 GND[K2]1.7
    GND[K2]J2 Open
    J3 GND[K3]1.7
    GND[K3]J3 Open
    J4 GND[K4]1.7
    GND[K4]J4 Open
    J5 GND[K5]1.7
    GND[K5]J5 Open
    J6 GND[K6]1.7
    GND[K6]J6 Open
    J7 GND[K7]1.7
    GND[K7]J7 Open
    J8 GND[K8]1.7
    GND[K8]J8 Open
    Figure 3. System Connector
  2. If one or more of the readings indicate a short, replace the coil. If all of the above readings are correct, redo Coil Imaging Performance Verification. If the problem still exists, replace the 1.5T HD 8 Channel High Resolution Brain Array Coil.

Topic ID: id_SL2763938-584045

Mechanical Parts Wear

Inspect the coil for visible signs of wear. If the coil housings do not seat with each other or if the latch mechanism will not work properly, contact GE Service.