• Discovery MR750 3.0T System Service Methods
  • 5690009-2EN Revision 4
  • Object ID: 00000018WIA30B30E20GYZ
  • Topic ID: id_13106628 Version: 3.0
  • Date: Aug 29, 2019 2:01:33 AM

3.0T 3-channel shoulder coil P port introduction

The performance of any modifications and/or repairs that have not been evaluated and approved by GEHC Coils Engineering may cause the device to be adulterated within the meaning of section 501(h) of the Food, Drug and Cosmetic Act, 21 U.S.C. § 351(h) and is strictly prohibited.

Specifically, unapproved modifications may impact the safety claims of the device pertaining to the patient/user in relation to shock, biocompatibility, flammability, biological, and thermal regulations. Any changes to materials outside of those within the products service documentation may be considered adulterated and can be in violation of good manufacturing practice requirements.

Product Identification and Shipping List

To identify the Shoulder coil, see the illustrations below.Figure 1.

Figure 1. Coil Label

Figure 2 shows a picture of the 3.0T Shoulder Coil.

Figure 2. 3.0T Shoulder Coil
Note:

Please note that the parts in Table 1 are not FRU. For FRU, please refer to 3.0T 3ch Shoulder Coil FRUs and Cable Replacement.

Table 1. Shipping List
Description Part Number Quantity
3.0T Shoulder Coil24143311
Shoulder Coil Patient Pad24235141
Shoulder Coil Spacer Pad24235161
Phantom Holder56943451
Shoulder Coil Strap Set53683901
Operators Guide 2422148-5001
3.0T HD SHOULDER ARRAY TRANSLATION LABEL5490496-111
Coil Name Language Translation Label Instruction Sheet54926581

Environmental Requirements

Storage Requirements

Coil should be stored in the scanner room.

Dimensions: L 217 mm (8.5 in) x W 232 mm (9.1 in) x H 316 mm (12.4 in)

Weight: 2.3 kg (5.0 lb)