• Topic ID: id_12373720
  • Version: 4.5
  • Date: Sep 17, 2019 3:57:22 PM

Checking and filling hydraulic fluid

Checks and adjusts the hydraulic fluid level for the dockable patient table.

Prerequisites

Personnel requirements
Required persons Preliminary requirements Procedure Finalization
1 - 30 minutes -
Tools and test equipment
Item Quantity Part number Manufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set 1 5113258 -
Consumables
Item Quantity Part number Manufacturer
Hydraulic Oil 1 5446125-2 -
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.