Removing the dockable table motor and clutch assembly

Removes the motor and ElectroMagnetic (EM) clutch assembly from the dockable table tabletop.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
The left- and right-hand table side covers are removed.

Procedure

  1. Undock the table and move it out of the scan room.
  2. Disconnect the motor power and signal connectors from the table wiring.
    Figure 1. Motor power and signal connectors
  3. Remove the cable ties securing the motor connectors to the table.
  4. Disconnect the EM clutch cable from the table wiring.
    Figure 2. EM clutch connector
  5. Raise the table handle.
  6. Loosen the screw securing the coupler to the rigibelt drive shaft.
    Note: If the coupler screw is not accessible, press and hold the dock pedal and move the cradle until the screw is accessible.
    Figure 3. Motor coupler screw
  7. Remove the four screws securing the motor bracket to the tabletop.
    Figure 4. Motor bracket screws
  8. Remove the motor and EM clutch assembly from the tabletop.
    Note: There are two small shaft keys in the coupler. Make sure the keys do not fall out of place during removal.