Set up the total digital imaging (TDI) of the lower posterior array (PA) for the Multi-Coil Quality Assurance (MCQA) check.
Procedure
- Put the flat filler panel closest to the bore on the table.
Figure 1. Flat filler panel
- Press the Alignment Light button.
- Move the cradle so the alignment light shows on the TDI PA coil lower end line on both sides of the cradle.
Note: There are lines on both sides of the cradle.
Figure 2. Flat filler panel alignment
- Do the following steps:
| Notice |
|---|
| Phantom spacer must be placed between two phantoms, otherwise the test will fail. |
- Place the two TL unified phantoms and the spacer on the table.
- Place PA phantom positioner over the two TL unified phantoms.
Figure 3. Phantom setup
- Move the phantom with positioner so that alignment line and alignment dent of the positioner are aligned with the beams.
Figure 4. Phantom alignment
- Press Landmark and then press Advance to Scan.
- Run the MCQA check.
Note: Select 3.0T_32Ch_TDI_PA_Lower (or equivalent platform option) in the Current Coil field.
Figure 5. MCQA tool menu