Setting up Total Digital Imaging (TDI) Posterior Array (PA)

Set up the Total Digital Imaging (TDI) of the upper and lower Posterior Array (PA) for the Multi-Coil Quality Assurance (MCQA) check.

Prerequisites

The 3.0T coil MCQA uses the 3.0T phantoms, which are pink in color.
Note: Coils do not ship with phantoms. Phantoms come in a unified phantom set with the system.
Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-60 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
TL Unified Phantom (for 3.0T)25343347-2-
PA Phantom Positioner for Both 1.5T and 3.0T15405748-
Phantom Positioner Spacer15628216-
Required conditions
Make sure dual drive quadrature calibration has been performed. Refer to Calibrating for quadrature.
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

  • Do the MCQA test for the TDI PA lower configuration.
  • Do the MCQA test for the TDI PA upper configuration.