Installs the oil catch pan onto the dockable patient table.
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
5 minutes
-
Procedure
From the bore-end of the table, slide the oil catch pan onto the rails under the table.
Note: Make sure the catch pan is oriented the correct direction as shown. An incorrectly installed catch pan may prevent the docking hook from engaging completely.
Figure 1. Oil catch pan
1
Small gap
2
Large gap
Press the caster lock pedal.
Figure 2. Brake caster lock pedal
Make sure that each wheel is locked by trying to rotate the table left and right from each end.
Result
No wheel should roll.
Note: Slight wheel caster from side-to-side on the rear wheels is allowed.
Press the free rotate pedal. Rotate the table left and right. Make sure that each wheel casters and rolls.
Figure 3. Brake free rotate pedal
Press the steer lock pedal. Rotate the table left and right. Make sure that both the rear and the right front wheels caster.
Figure 4. Brake steer lock pedal
Result
The left front wheel should remain oriented longitudinally along the table (it could rotate up to 180 degrees before locking).
If any of the above brake checks fail, make sure the oil catch pan is not interfering with brake linkages.