• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30B56C40GYZ
  • Topic ID: id_2019479 Version: 3.1
  • Date: Jul 13, 2021 4:10:15 PM

Removing the dockable table Physiological Acquisition Controller (PAC)

Removes the Physiological Acquisition Controller (PAC) from the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
The right-hand base side cover has been removed.
The table is disconnected or ISC LOTO is applied.

Procedure

  1. If it is not already undocked, undock the table and move it away from the magnet.
    Important: Undocking the table disconnects the power to the Physiological Acquisition Controller (PAC). If the table is docked during this procedure, ISC LOTO is required.
  2. Disconnect the following cables and pneumatic tube from the PAC:
    Figure 1. PAC connections
    1ECG extension cablePAC ECG connector
    2M3386J6
    3M3386J5
    45946424J2
    5Resp tubeJ3
    65946423J1
    75946424J9
  3. Remove the four screws and four washers securing the PAC to the table.
    Figure 2. PAC hardware
    1Washer
    2Screw
  4. Cut cable ties securing cables to the PAC bracket.
  5. Remove the PAC from the table.