Installing the dockable table Physiological Acquisition Controller (PAC)

Installs the Physiological Acquisition Controller (PAC) into the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Consumables
ItemQuantityPart numberManufacturer
Cable Tie, Self-LockingAs needed5343293-

Procedure

  1. Install the PAC into the table.
  2. Loosely install the four screws securing the PAC to the table.
    Figure 1. PAC hardware
    1Washer
    2Screw
  3. Connect the following cables and pneumatic tube to the PAC:
    Figure 2. PAC connections
    1ECG extension cablePAC ECG connector
    2M3386J6
    3M3386J5
    45946424J2
    5Resp tubeJ3
    65946423J1
    75946424J9
  4. Tighten the four screws securing the PAC to the table.
  5. Install cable ties securing the cables and pneumatic tube to the PAC bracket and table cable track as required.
    Note: Do not overtighten the cable ties.
  6. Lower the table to the down position to make sure there is no interference with the cable routing.
  7. Re-route any cables that are touching table structure.