Installs the Physiological Acquisition Controller (PAC) into the dockable table.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 15 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Consumables |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Cable Tie, Self-Locking | As needed | 5343293 | - |
Procedure
- Install the PAC into the table.
- Loosely install the four screws securing the PAC to the table.
Figure 1. PAC hardware
- Connect the following cables and pneumatic tube to the PAC:
Figure 2. PAC connections
| 1 | ECG extension cable | PAC ECG connector |
| 2 | M3386 | J6 |
| 3 | M3386 | J5 |
| 4 | 5946424 | J2 |
| 5 | Resp tube | J3 |
| 6 | 5946423 | J1 |
| 7 | 5946424 | J9 |
- Tighten the four screws securing the PAC to the table.
- Install cable ties securing the cables and pneumatic tube to the PAC bracket and table cable track as required.
Note: Do not overtighten the cable ties.
- Lower the table to the down position to make sure there is no interference with the cable routing.
- Re-route any cables that are touching table structure.