Checking the table undock mechanism components

Checks the operation of the table undock mechanism components and primary latch adjustment.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-

Procedure

  1. Move the table into the scan room and dock it.
  2. Advance the cradle into the bore until the primary latch is exposed.
  3. Move the cradle back onto the table.
  4. Undock the table.
  5. If the cradle fails to move fully off of or onto the table, adjust the cradle home table down interlock. Refer to Adjusting the cradle home table down interlock.